Automated blood sampling systems for positron emission tomography

Lars I. Eriksson, S. Holte, C. Böhm, M. Kesselberg, B. Hovander

IEEE Transactions on Nuclear Science · 1988 · 140 citations · 2 references

Concepts

Abstract

An automated blood sampling system has been constructed and evaluated for use with positron-emission tomography (PET). The dispersion functions of two different detector units in the blood sampling system are compared. One is a plastic scintillator, and the other is a coincidence detector. No significant differences were found. Results from studies of blood-brain barrier transfer of a C-11 labelled receptor antagonist are discussed.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>