MedicineRadiologyDiagnostic ImagingUltrasonographyUltrasoundCalibrationDiagnostic UltrasonographyTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound InstrumentationB-mode ImagingB-mode UltrasonographyCalibration ErrorPrecisionSurface RegistrationTransducer CalibrationUltrasound CalibrationUltrasound Probe Calibration
OBJECTIVE: Precise transducer calibration is an essential prerequisite for reliable surface registration based on ultrasound B-mode imaging devices. The clinical usage of a novel B-mode transducer calibration technique was evaluated and its attainable calibration precision assessed. MATERIALS AND METHODS: The Three Wire Method and the Cambridge Calibration Method were used as reference techniques to compare the efficiency, calibration precision and spatial requirements of the different techniques. A total of 20 calibration trials were performed using each technique and were statistically evaluated for accuracy and speed. RESULTS: The mean error characterizing the calibration precision of the Three Wire Method was 3.2 mm, obtained in a phantom with a volume of 14 x 10(6) mm(3) in 18.48 min. The Cambridge method resulted in a mean calibration error of 2.2 mm, but required a larger phantom with a volume of 35 x 10(6) mm3 to be used for a duration of 9.30 min. The proposed method yielded an average calibration error of 1.9 mm and was performed, on average, in 2 min using a phantom with a size smaller than 1 x 10(6) mm3. CONCLUSIONS: The suggested calibration method offers decreased time and space while retaining an equivalent calibration precision when compared to established reference methods.
7
Rapid calibration for 3-D freehand ultrasound
Richard W. Prager, Robert Rohling, Andrew H. Gee et al. · Ultrasound in Medicine & Biology · 1998 · 386 citations
Three-dimensional freehand ultrasound: Image reconstruction and volume analysis
Christin Barry, C Allott, Nigel W. John et al. · Ultrasound in Medicine & Biology · 1997 · 198 citations