Gas chromatography/combustion/isotope‐ratio mass spectrometry analysis of urinary steroids to detect misuse of testosterone in sport

Michel Becchi, Rodrigue Aguilera, Yves Farizon, Marie‐Magdeleine Flament, Hervé Casabianca, Patrick James

Rapid Communications in Mass Spectrometry · 1994 · 183 citations · 8 references

Concepts

Abstract

We propose a new confirmatory method for testosterone doping in sport. The present method in use, based on measuring the testosterone/epitestosterone (T/E) ratio in urine, may miss suspicious cases, or lead to reporting cases in which the high ratio is natural. Synthetic testosterone has a 13C abundance different from that of endogenous human testosterone. The connection of a gas chromatograph to an isotope-ratio mass spectrometer via a combustion interface allows the measurement of the corresponding characteristic value (delta /1000) for testosterone, its precursors, and its metabolites. To detect exogenous administration of testosterone, 30-40 mL of urine is sufficient.

References

8