Journal of Magnetic Resonance Imaging · 1994 · 58 citations · 16 references
MedicineCardiologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyFerromagnetismMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic SusceptibilityMedical ImagingEx Vivo ImagingFerromagnetic ResonanceMagnetic ResonanceMagnetic Resonance Imaging AcquisitionMagnetic Resonance PhysicsProsthetic Heart Valve ImagingSafety In Magnetic Resonance ImagingStatic Magnetic Field
Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5-T (64-MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5-T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient-recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a "worst case" MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2 degree C with the implant imaged "in air" and +0.3 degrees C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5-T (64-MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.