Testing a new NIF neutron time-of-flight detector with a bibenzyl scintillator on OMEGA

V. Yu. Glebov, C. J. Forrest, J. P. Knauer, A. Pruyne, M. H. Romanofsky, T. C. Sangster, M. J. Shoup, C. Stöeckl, J. A. Caggiano, M. Leslie Carman,

Review of Scientific Instruments · 2012 · 36 citations · 8 references

Concepts

Abstract

A new neutron time-of-flight (nTOF) detector with a bibenzyl crystal as a scintillator has been designed and manufactured for the National Ignition Facility (NIF). This detector will replace a nTOF20-Spec detector with an oxygenated xylene scintillator currently operational on the NIF to improve the areal-density measurements. In addition to areal density, the bibenzyl detector will measure the D-D and D-T neutron yield and the ion temperature of indirect- and direct-drive-implosion experiments. The design of the bibenzyl detector and results of tests on the OMEGA Laser System are presented.

References

8