Obstetrics and Gynecology · 2015 · 63 citations · 23 references
MedicineSurgeryUrologyBowel ManagementFemale UrologyUrodynamicsContinence AssessmentFecal IncontinenceFecal Incontinence ManagementIncontinence ManagementPelvic Floor DisordersUrinary IncontinenceBowel ControlExtended-wear TherapeuticsIncontinence Episode FrequencyIncontinence Severity MeasurementIntravaginal TherapyPatient-reported Bowel SymptomsSymptom Score ImprovementTreatment Success Predictors
OBJECTIVE: To evaluate the effectiveness and safety of a vaginal bowel-control device and pump system for fecal incontinence treatment. METHODS: Women with a minimum of four fecal incontinence episodes over 2 weeks were fit with the intravaginal device. Treatment success, defined as a 50% or greater reduction of incontinent episodes, was assessed at 1 month. Participants were invited into an optional extended-wear period of another 2 months. Secondary outcomes included symptom improvement measured by the Fecal Incontinence Quality of Life, Modified Manchester Health Questionnaire, and Patient Global Impression of Improvement. Adverse events were collected. Intention-to-treat analysis included participants who were successfully fit entering treatment. Per protocol, analysis included participants with a valid 1-month treatment diary. RESULTS: Sixty-one of 110 (55.5%) participants from six clinical sites were successfully fit and entered treatment. At 1 month, intention-to-treat success was 78.7% (48/61, P<.001); per protocol success, 85.7% (48/56, P<.001) and 85.7% (48/56) considered bowel symptoms "very much better" or "much better." There was significant improvement in all Fecal Incontinence Quality of Life (P<.001) and Modified Manchester (P≤.007) subscales. Success rate at 3 months was 86.4% (38/44; 95% confidence interval 73-95%). There were no serious adverse events; the most common study-wide device-related adverse event was pelvic cramping or discomfort (25/110 participants [22.7%]), the majority of events (16/25 [64%]) occurring during the fitting period. CONCLUSION: In women successfully fit with a vaginal bowel-control device for nonsurgical treatment for fecal incontinence, there was significant improvement in fecal incontinence by objective and subjective measures. CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov, www.clinicaltrials.gov, NCT01655498. LEVEL OF EVIDENCE: : II.
23
Fecal incontinence quality of life scale
Todd Rockwood, James M. Church, James W. Fleshman et al. · Diseases of the Colon & Rectum · 2000 · 1.2K citations
Sacral Nerve Stimulation for Fecal Incontinence
Steven D. Wexner, John A. Coller, Ghislain Devroede et al. · Annals of Surgery · 2010 · 350 citations