Magnetic Resonance in Medicine · 1999 · 91 citations · 24 references
Natural SciencesRadiologyDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance BiophysicsImagingMagnetic Resonance Contrast MechanismsMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingArtifacts In Magnetic Resonance ImagingEcho Time DependenceMagnetic ImagingMagnetic Resonance Imaging AcquisitionTime-varying Magnetic Fields
Maxwell's equations imply that imaging gradients are accompanied by higher order spatially varying fields (concomitant fields) that can cause artifacts in MR imaging. The lowest order concomitant fields depend quadratically on the imaging gradient amplitude and inversely on the static field strength. Time-varying concomitant fields that accompany the readout gradients of spiral scans cause unwanted phase accumulation during the readout, resulting in spatially dependent blurring. Concomitant field phase errors are independent of echo time and, therefore, cannot be detected using Dixon-type field map measurements that are normally used to deblur spiral scan images. Data acquisition methods that reduce concomitant field blurring increase off-resonant spin blurring, and vice versa. Blurring caused by concomitant fields can be removed by variations of image reconstruction methods developed to correct for spatially dependent resonance offsets with nonrectangular k-space trajectories.
24
Fast Spiral Coronary Artery Imaging
Craig H. Meyer, Bob S. Hu, Dwight G. Nishimura et al. · Magnetic Resonance in Medicine · 1992 · 885 citations
Medicine, Cardiology, Radiology +14