Applied Physics Letters · 2008 · 36 citations · 13 references
Natural SciencesNuclear EngineeringPhysicsNuclear InstrumentationNuclear PhysicsRadiation DetectionAlpha DecayAlpha Particle DetectionNuclear DecayNuclear Material MeasurementParticle DetectorsRadioactive DecayAlpha Particle DetectorsAlpha Particle SpectrometryAlpha SpectrometryAlpha SpectroscopyAlpha-particle SpectroscopyDetectorsDirect MeasurementMicrocalorimeter Detectors
Identification of trace nuclear materials is usually accomplished by alpha spectrometry. Current detectors cannot distinguish critical elements and isotopes. We have developed a detector called a microcalorimeter, which achieves a resolution of 1.06 keV for 5.3 MeV alphas, the highest resolving power of any energy dispersive measurement. With this exquisite resolution, we can unambiguously identify the P240u/P239u ratio in Pu, a critical measurement for ascertaining the intended use of nuclear material. Furthermore, we have made a direct measurement of the P209o ground state decay.
13
Single optical photon detection with a superconducting tunnel junction
A. Peacock, P. Verhoeve, N. Rando et al. · Nature · 1996 · 230 citations
The world of physical chemistry
Choice Reviews Online · 1994 · 154 citations
Nuclear data sheets for A = 209
M.J. Martin · Nuclear Data Sheets · 1991 · 130 citations