Hip Screw Augmentation with an In Situ-Setting Calcium Phosphate Cement: An In Vitro Biomechanical Analysis

69

Citations

19

References

1997

Year

Abstract

The in situ-setting calcium phosphate cement investigated in this study compared favorably with PMMA in a single-cycle cut-out test of augmented compression hip screws in senile trabecular bone. Our results suggest that these materials may have promise as substitutes for PMMA in the salvage of compression hip screw fixation in elderly osteopenic patients with complex intertrochanteric fractures and that further study of their use in this application is warranted.