Severe Methemoglobinemia Detected by Pulse Oximetry

Emil Annabi, Steven J. Barker

Anesthesia & Analgesia · 2009 · 46 citations · 8 references

Concepts

Abstract

In Brief An elderly surgical patient acquired a life-threatening methemoglobinemia as a result of topical benzocaine spray to the oropharynx in preparation for awake endotracheal intubation. A new multiwavelength pulse oximeter, the Masimo Rad-57, detected this methemoglobinemia an hour before it was confirmed by laboratory CO-oximetry. The Rad-57 monitored the patient's methemoglobin levels during diagnosis and treatment with methylene blue, and the values it provided (as high as 33%) were very close to those of the laboratory CO-oximeter. The new pulse oximeter gave continuous readings of methemoglobin level at the bedside, whereas the laboratory values were delayed by up to an hour. This case demonstrates the clinical application of a multiwavelength pulse oximeter in the diagnosis and treatment of a life-threatening dyshemoglobinemia. IMPLICATIONS: This case report demonstrates that life-threatening, drug-induced methemoglobinemia can be diagnosed and monitored using a new multiwavelength pulse oximeter, the Masimo Rad-57. This new device can detect methemoglobin without the delays inherent in laboratory blood analysis, and it can monitor the effectiveness of treatment.