MedicineOtolaryngologyRadiologyCochlear ImplantsDiagnostic ImagingDiagnostic RadiologyCochlear DisordersCochlear ImplantationImplant TechnologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsSurgical ImplantationCochlear Implant CompatibilityElectromagnetic InterferenceElectromagnetic Interference MitigationElectromagnetic ShieldingImaging For Cochlear ImplantationMagnetic ResonanceMagnetic Resonance PhysicsMedical Indication
In vitro experiments were performed to determine the compatibility of a cochlear implant at 0.2- and 1.5-T magnetic resonance (MR) imaging. Except for the torque at 1.5 T, all electromagnetic interferences remained within acceptable limits. MR imaging should be performed only if there is a strong medical indication, by following appropriate safety procedures. MR imaging at 0.2 T should be safe; at 1.5 T, however, the relative risks will have to be assessed.