The Analyst · 1984 · 52 citations · 0 references
PharmacyPharmaceutical AnalysisAnalytical ChemistryClinical BiochemistryElemental AnalysisAnalytical CalibrationAnalytical MeasurementsBioanalytical Method ValidationHeavy Metal AnalysisInstrumental AnalysisTrace Element AnalysisAtomic Emission SpectrometryBiomedical Elemental AnalysisBiomedical Sample AnalysisInductively Coupled Plasma Atomic Emission SpectroscopyInductively Coupled Plasma SpectrometryInterferenceSample InjectionSerum Magnesium Measurement
A rapid and precise method for the simultaneous determination of Na, K, Ca, Mg, Li, Cu, Fe, and Zn in blood serum by flow injection analysis-inductively coupled plasma atomic-emission spectrometry (FIA-ICP-AES) has been developed using aqueous synthetic multi-element solutions for calibration. Performance characteristics of the FIA-ICP-AES technique are documented in terms of the effects of sample injection volume and carrier stream flow-rate on sensitivity and precision. Matrix interference was detected for injections of undiluted serum and the effect was minimised at low injection volumes and/or by using a relatively high RF power. At dilution factors of 1 + 1 (or greater) interference was not observed and reliable data were obtained for human and bovine serum pools for sample injections of 20 µl.