Journal of Neural Engineering · 2010 · 86 citations · 35 references
EngineeringBiomedical EngineeringNeuroscienceElectrophysiologyNeurophysiologyProstheticsElectrode Array DesignElectrode PlacementEvoked PotentialsImplantable Biomedical DevicesNeural ProstheticsNeurostimulationCortical Evoked ResponsesElectrical StimulationElectrode GeometryElectrode Geometry EffectsMonopolar StimulationRetinal ProsthesesRetinal ProstheticsStimulus Parameters
Several approaches have been proposed for placement of retinal prostheses: epiretinal, subretinal and suprachoroidal. We aimed to systematically evaluate the effectiveness of varying a range of stimulus parameters and electrode geometry for a suprachoroidal electrode array, using cortical evoked responses to monopolar electrical stimulation in cats. Our results indicate that charge thresholds were not dependent on electrode size, pulse widths or position of the return electrode tested, but were dependent on the number of sites stimulated in parallel. Further, we found that the combination of monopolar stimulation with large diameter electrodes, wide pulse widths and parallel stimulation minimized the voltage requirements for stimulation. These results provide useful insights for the design specifications of a low voltage suprachoroidal stimulator.
35
Better speech recognition with cochlear implants
Blake S. Wilson, Charles C. Finley, Dewey T. Lawson et al. · Nature · 1991 · 1.2K citations
Health Professions, Otolaryngology, Speech-language Pathology +14
Preservation of the Inner Retina in Retinitis Pigmentosa
Arturo Santos · Archives of Ophthalmology · 1997 · 643 citations