Differences in effective dose estimation from dose–area product and entrance surface dose measurements in intravenous urography
53
Citations
11
References
2001
Year
Dose–area ProductRadiation ExposureTreatment VerificationIntravenous UrographyUrogenital RadiologyEntrance Surface DoseRadiation Therapy PlanningPediatric DosimetryRadiation RiskInstrumentationEffective Dose EstimationRadiation OncologyNuclear MedicineRadiologyHealth SciencesMedical ImagingRadiographic ImagingDosimetryUrologyRadiation DoseMedicineDose-area Product
In this study, measurements of dose-area product (DAP) and entrance surface dose (ESD) were carried out in a sample of 25 adult patients who underwent intravenous urography (IVU). These measured quantities were used to estimate the effective dose E from the IVU examination, a quantity closely correlated to radiation risk. Estimating E involves the use of conversion coefficients that have been determined for specific X-ray views in a mathematical phantom. These are obtained under conditions which are not usually met in clinical practice. As a result, the E estimates using the two different measurable quantities can be quite different. Analysis of the calculation procedure suggests that the E estimate using the DAP measurements, in addition to being more practical, could be more accurate than using ESD measurements, as DAP is sensitive to the X-ray field size settings. Furthermore, it is shown that in the absence of the appropriate equipment, a reliable E estimate can be obtained from the ESD calculated using the exposure data for each X-ray view.