Vitreous flow rates through dual pneumatic cutters: effects of duty cycle and cut rate

Dina Joy K. Abulon

Clinical ophthalmology · 2015 · 32 citations · 7 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: We aimed to investigate effects of instrument settings on porcine vitreous flow rates through dual pneumatic high-speed vitrectomy probes. METHODS: The CONSTELLATION(®) Vision System was tested with 250, 450, and 650 mmHg of vacuum using six ULTRAVIT(®) vitrectomy probes of each diameter (25+(®), 25, 23, and 20 gauge) operated from 500 cuts per minute (cpm) up to 5,000 cpm. Duty cycle modes tested included biased open, 50/50, and biased closed. Flow rates were calculated by assessing the change in weight of porcine eyes during vitreous aspiration. Volumetric flow rate was measured with a computer-connected electronic scale. RESULTS: At lower cut rates, the biased open mode produced higher flow than did the 50/50 mode, which produced higher flow than did the biased closed mode. In the biased closed and 50/50 modes, vitreous flow rates tended to increase with increasing cut rate. Vitreous flow rates in the biased open duty cycle mode remained relatively constant across cut rates. CONCLUSION: Vitreous flow rates through dual pneumatic vitrectomy probes could be manipulated by changing the duty cycle modes on the vitrectomy system. Differences in duty cycle behavior suggest that high-speed cut rates of 5,000 cpm may optimize vitreous aspiration.