Journal of High Resolution Chromatography · 1986 · 24 citations · 8 references
Natural SciencesChemistryEnvironmental SciencesAnalytical ChemistryEnvironmental ChemistryMass Spectrometric AnalysisChromatographyEnvironmental AnalysisEnvironmental Analytical ChemistryGas ChromatographyInstrumental AnalysisMass SpectrometryAnalytical Detection Of Organic PollutantsMass Spectrometric IdentificationOpen Tubular ColumnsPolychlorinated Biphenyl CharacterizationPolychlorinated BiphenylsQuantitative Gas ChromatographySelected Ion MonitoringSpeciation In Environmental Samples
Abstract An integrated analytical procedure for polychlorinated biphenyl (PCB) multiresidues in environmental samples has been developed and applied to well‐characterized samples. PCB residues are characterized and quantified using wall coated open tubular column (WCOT) gas chromatography or HRGC/MS‐selected ion monitoring. An automated data system based on a Spectra Physics SP‐4200 computing integrator is used to select and quantitate peaks of individual congeners and homolog groups. This procedure consistently yields results with a reproducibility within 3 percent for HRGC‐ECD and 3.7 percent for HRGC/MS‐SIM technique respectively.
8
L.G.M.Th. Tuinstra, A.H. Roos, B. Griepink et al. · Journal of High Resolution Chromatography · 1985 · 40 citations
Pharmacy, Chemistry, Toxicology +17