Analytical Chemistry · 2014 · 172 citations · 16 references
Life SciencesBiotechnologyMolecular BiologyBiosensorsMolecular DiagnosticsVirologyDna BiosensorsLateral Flow ImmunoassaysMultiplexed BiosensingNucleic Acid BiosensorsPolymerase Chain ReactionViral DiagnosticsAssayAssay SpecificityFluorescent Dna BiosensorsFluorescent Nucleic Acid ProbesFluorescent ProbesLateral Flow AssaysMultiplex Lateral Flow AssaysMultiplex Nucleic Acid Detection
Here we report a rapid, low cost, and disposable dipstick-type DNA biosensor that enables multiplex detection in a single assay. The fluorescent probes labeled with different fluorophores were introduced into the lateral flow nucleic acid testing system. In combination with multiple immobilized probes arranged in an array formant on the membrane, a dual-color fluorescent lateral flow DNA biosensor was developed using a portable fluorescence reader. Up to 13 human papillomavirus types could be detected simultaneously by a single-step operation in less than 30 min after linear-after-the-exponential (LATE)-PCR. The sensitivity was determined to be 10-10(2) copies plasmid DNA/μL. The specificity study showed no cross-reactivity among the 31 different common HPV types. In the clinical validation, 95.3% overall agreement showed very good potential for this method in the clinical application when compared to a commercial kit.
16
Ana Maria de Roda Husman, Jan M.M. Walboomers, A. J. C. van den Brule et al. · Journal of General Virology · 1995 · 1.3K citations
J. Aquiles Sanchez, Kenneth E. Pierce, John E. Rice et al. · Proceedings of the National Academy of Sciences · 2004 · 235 citations · Full text