IEEE Transactions on Biomedical Engineering · 1984 · 92 citations · 8 references
Veterinary MedicineBiomedical EngineeringCardiologyDiagnostic ImagingUltrasonographyUltrasoundCardiac ImagingCardiac UltrasonographyCardiac UltrasoundCardiovascular ImagingMedical Image AnalysisMedical ImagingClinical Validation StudiesDiagnostic Validation StudiesIntracardiac EchocardiographySignal-to-noise RatioSmoothingSpatio-temporal SmoothingTemporal SmoothingTwo-dimensional Echocardiography
An effective image processing algorithm for extracting the endocardial border from a series of 2-dimensional echocardiograms (2D echo) is proposed. The main steps involved in this algorithm are coordinate transformation, regional co-occurrence matrix thresholding, search region limiting, border detecting, and spatial and temporal smoothing. Using these steps, some major difficulties related to 2D echo image processing including low signal-to-noise ratio, echo drop out, and gray level intensity variability have been dramatically reduced. All endocardial borders in an entire cardiac cycle are computed. This is done automatically on the basis of observer-drawn outlines at end systole and end diastole. These outlines facilitate the quantitative analysis by defining a range in which both boundary dimension and shape changes should occur. Image processing is restricted to this range. The validation studies have been perfonned on test balloons where both image quality has been changed and complex motion abnormalities have been experimentally produced. Good results have also been obtained in patient studies.
8
Edward D. Folland, Attilio Parisi, P F Moynihan et al. · Circulation · 1979 · 729 citations · Full text
Medicine, Cardiology, Radiology +17