Photoplethysmography imaging: a new noninvasive and noncontact method for mapping of the dermal perfusion changes

Ting Wu, Vladimír Blažek, Hans J. Schmitt

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2000 · 117 citations · 0 references

Concepts

Abstract

This paper presents the experimental setup and preliminary results of a CCD based Photoplethysmographic Imager (PPGI) which has been shown to be capable to assess various disorders of the peripheral venous system by standard test methods derived from the classical photophethysmographic practice in a noninvasive and non-contact way. The PPGI is a computer- based CCD imaging system to visualize the skin vessels and analyze the local changes of dermal blood volume. Our results show that this system performs as well as the currently available commercial PPG system by adding information of spatial distribution which allows the investigation of locations and causes of vascular disorders. Both the venous hemodynamics and arterial perfusion can be mapped in two dimensions.