Journal of Clinical Microbiology · 2002 · 72 citations · 12 references
Life SciencesInfectious DiseasesMicrobiologyDiagnostic MicrobiologyMedical MicrobiologyVirologyInfectious Disease DetectionInfectious Disease Molecular TestingPathogen DetectionPolymerase Chain ReactionRapid Diagnostic TestsViral DiagnosticsGenital UlcersHerpes Simplex Virus Infection DiagnosisHerpes Simplex Virus TestingPcr-based DetectionRapid Detection AssaysRapid Viral DiagnosticsViral Genome DetectionVirus Detection Assays
We have evaluated a real-time PCR procedure based on the LightCycler technology for rapid detection of herpes simplex virus (HSV) in genital lesions. Two sets of primers, corresponding to the thymidine kinase and DNA polymerase regions, were used for the amplification reactions in separate capillaries containing the SYBR Green I dye as detection signal. In 28 of 118 samples (24%), HSV was isolated by conventional cell culture. All cell culture-positive samples were also positive by real-time PCR. Six additional cell culture-negative samples were positive by PCR with both sets of primers. Total processing time was less than 3 h. Real-time PCR using SYBR Green I as detection signal is a sensitive procedure for the rapid diagnosis of HSV in genital lesions.