Publication | Open Access
Validation of a fully autonomous phosphate analyser based on a microfluidic lab-on-a-chip
36
Citations
13
References
2010
Year
EngineeringAvailable Phosphate MonitorAnalytical MicrosystemsBench Top CalibrationBiomedical EngineeringBiosensing SystemsBioanalysisBioprocess MonitoringAnalytical ChemistryAnalytical BiotechnologyMicrofluidicsAutonomous Phosphate AnalyserChromatographyMicrofluidic Lab-on-a-chipBiomedical AnalysisPhosphate AnalyserWaste ManagementWater AnalysisEnvironmental EngineeringMass SpectrometryLab-on-a-chipBiomemsMedicine
This work describes the design of a phosphate analyser that utilises a microfluidic lab-on-a-chip. The analyser contains all the required chemical storage, pumping and electronic components to carry out a complete phosphate assay. The system is self-calibrating and self-cleaning, thus capable of long-term operation. This was proven by a bench top calibration of the analyser using standard solutions and also by comparing the analyser's performance to a commercially available phosphate monitor installed at a waste water treatment plant. The output of the microfluidic lab-on-a-chip analyser was shown to have sensitivity and linear range equivalent to the commercially available monitor and also the ability to operate over an extended period of time.