Thermal positive-ion source developed for molecular beam detection

Hiroyuki Kawano, Seiji Matsui, Naoshi Serizawa

Review of Scientific Instruments · 1996 · 14 citations · 4 references

Concepts

Abstract

A thermal positive-ion source developed for neutral beam detection has a filament (∼0.1 mm in diameter) of Re or W, which is surrounded with a cylinder. A beam of metal halide (MX) was directed onto a small area (S≊10−3 cm2) of the filament at a constant flux (N), and M+ emitted from the filament was collected mainly with the cylinder and partly (∼1%) with a Faraday cup. The ionization efficiency of both InI and TlI on Re in high vacuum (∼0.2 μTorr), for example, is unity in the optimum temperature range of ∼1000–1100 K mainly because the work function of Re under the above condition is as large as ∼7 eV, thereby yielding the ion current of ∼1 nA at N≊1012 molecules cm−2 s−1. Our experimental data and analytical results indicate that Re as well as W in the optimum range detects efficiently (∼100%) a beam of MX impinging at a net flux (NS) of more than ∼1010 molecules s−1 and yields ∼1 nA of M+ so long as the ionization energy of M is less than ∼6.4 eV. The data on the range are summarized together with experimental conditions employed.