Publication | Closed Access
Transcutaneous bilirubin – comparing the accuracy of BiliChek<sup>®</sup>and JM 103<sup>®</sup>in a regional postnatal unit
20
Citations
17
References
2010
Year
Both TcB devices correlated closely with moderately increased TSB levels and are suitable screening tools to identify jaundiced infants that require a serum bilirubin, with upper limit cut-off values. Both devices reduced the need for TSB levels. We found the BiliChek® slightly more accurate than the JM 103® for our study population. TcB however, is not in widespread use.