Journal of Applied Crystallography · 1991 · 11 citations · 0 references
EngineeringRadiologyRadiation DetectionRadiation DosimetryRadiation MeasurementDead Time CorrectionX-ray DetectionX-ray DetectorsCount Rate AnalysisDead Time EffectsDead Time In Photon CountingDead Time In Radiation DetectorsDetector Dead TimeNon-extended Dead Time ModelingSystem Dead Time Characterization
A practical equation is presented to determine the deadtime τd in conventional X-ray counting systems when the amount of deadtime is greater than that which can be accounted for by non-extended deadtime theory. The use of this equation to determine the deadtime increases the practical range over which conventional X-ray detection systems can be used.