Effects of Charité Artificial Disc on the Implanted and Adjacent Spinal Segments Mechanics Using a Hybrid Testing Protocol

230

Citations

32

References

2005

Year

Abstract

The hybrid testing protocol is advocated because it better reproduces clinical observations in terms of motion following surgery, using pure moments. Using this approach, we found that the Charité artificial disc placement slightly increases motion at the implanted level, with a resultant increase in facet loading when compared to the adjacent segments, while the motions and loads decrease at the adjacent levels. However, in the load control mode that we believe is not that clinically relevant, there was a large increase in motion and a corresponding increase in facet loads, as compared to the intact.