Journal of Chromatographic Science · 1980 · 38 citations · 0 references
PharmacyChemical EngineeringChemistryChromatographic AnalysisPolymer ChemistryPolymer EngineeringChromatographyLiquid ChromatographyMobile Phase OptimizationParticle Size ControlStationary Phase ChemistrySupercritical Fluid ChromatographyColumn Length Effects On ResolutionFlow Rate DependenceParticle Size Effects On ResolutionPolystyrenePressure ProgrammingSupercritical Fluid Chromatography ColumnsThermophysical PropertiesTurbulent Flow Regime
In pressure-programmed supercritical fluid chromatography, particle diameters in the range of 122 μm to 205 μm were found to yield maximum resolution of oligomers of 800 molecular weight polystyrene when 5% isopropanol in pentane was employed as a mobile phase at 240°C. The ability of a given column to achieve maximum resolution was found to depend on the ratio of the pressure-programming rate to the average flow rate of the mobile phase. Resolution could be increased while analysis time decreased, when a fast pressure-programming rate was used in conjuction with a high flow rate. In addition, due to the unusual thermophysical properties of the super-critical mobile phase, operation at linear velocities that are generally considered to be in the region of turbulent flow were found to be necessary for maximum resolution.