European Journal of Echocardiography · 2006 · 10 citations · 16 references
MedicineCardiologySurgeryCardiothoracic SurgeryInterventional CardiologyPulmonologyPatent Foramen OvalePatent Foramen Ovale ClosurePericardial DiseasePericardial SurgerySeptal DefectsThrombosisAtrial Septal Defect EvaluationPatent Foramen Ovale AnatomyPercutaneous Atrial Septal Defect ClosurePercutaneous Patent Foramen Ovale ClosurePercutaneous Septal Defect ClosurePericardial Patch RepairSurgical Patent Foramen Ovale ClosureThrombotic Lesion Description
We are presenting a case of floating left and right atrial formations on an atrial septal defect occluder system (23mm StarFLEX)-Occluder) initially supposed to be thrombotic appositions in a 57-year-old man. The closure was performed on the background of left hemispheric stroke and atrial septal aneurysm (ASA) with patent foramen ovale (PFO). The suspect structures were detected in the 6-month follow-up by transesophageal echocardiography (TEE). The patient underwent a successful surgical explantation of the closure device and closure of the patent foramen ovale (PFO) using a pericardial patch. The pathological evaluation of the biatrial device associated appositions revealed hytrophic heart muscle tissue with perifocal scarring and purulent abscess-forming, granulating and foam-cell including inflammatory foreign body reaction instead of the expected thrombus formation.
16