A Unifying Parameter to Describe the Clinical Mechanics of Hydrogel Contact Lenses

Donald Martin, JOHN BOULOS, JAMES GAN, KYPROS GAVRIEL, Paul Harvey

Optometry and Vision Science · 1989 · 35 citations · 0 references

Concepts

Abstract

We studied the relation of postlens squeeze pressure to the clinical fit of hydrogel contact lenses. This relation was statistically significant (p less than 0.01) for a variety of contact lenses currently prescribed over a range of thickness, water content, back central optic radius, diameter, and back vertex power. The squeeze pressure provided a useful parameter to describe and compare the clinical fit of hydrogel contact lenses and a model with which the fit of a lens to an eye could be predicted. There was virtually no movement of the contact lenses across the eye when the squeeze pressure was more than -14 mm H2O (-1370 dynes/cm2). However, the mechanics of the lens motion when the squeeze pressure was less than this critical pressure appeared to be related to a more complicated combination of the squeeze pressure and the other forces in the eye-lens system.