Phantomless patient‐specific TomoTherapy QA via delivery performance monitoring and a secondary Monte Carlo dose calculation
42
Citations
25
References
2014
Year
MCLogQA is a new TomoTherapy QA process that validates the planned dose before delivery and analyzes the delivered dose using the treatment exit detector and log file data. The MCLogQA procedure is an effective and efficient alternative to traditional phantom-based TomoTherapy plan-specific QA because it allows for comprehensive 3D dose verification, accounts for tissue heterogeneity, uses patient CT density tables, reduces total QA time, and provides for a comprehensive QA methodology for each treatment fraction.