Current Thyroglobulin Autoantibody (TgAb) Assays Often Fail to Detect Interfering TgAb that Can Result in the Reporting of Falsely Low/Undetectable Serum Tg IMA Values for Patients with Differentiated Thyroid Cancer
166
Citations
38
References
2011
Year
Many specimens with interfering TgAb were misclassified as TgAb negative using manufacturer-recommended cutoffs. It is recommended that assay AS limits be used to detect TgAb to minimize false-negative misclassifications. However, for two of four assays, AS limits failed to detect interfering TgAb in 20-30% of cases. TgAb methods were too qualitatively and quantitatively variable to establish conversion factors that would allow a change in method without disrupting serial TgAb monitoring.