Analytical Chemistry · 2004 · 310 citations · 9 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsMolecular RecognitionSensorsBiological SensorsDna BiosensorsMicroscale BiosensorsMolecular SensorsNucleic Acid BiosensorsPolymerase Chain ReactionAptamer-based BiosensingAptamer-based SensorsAptamersCantilever Beam BendingCantilever MechanicsDna Aptamer-based BiosensorsLabel-free DetectionSurface Stress Sensing
We report label-free protein detection using a microfabricated cantilever-based sensor that is functionalized with DNA aptamers to act as receptor molecules. The sensor utilizes two adjacent cantilevers that constitute a sensor/reference pair and allows direct detection of the differential bending between the two cantilevers. One cantilever is functionalized with aptamers selected for Taq DNA polymerase while the other is blocked with single-stranded DNA. We have found that the polymerase-aptamer binding induces a change in surface stress, which causes a differential cantilever bending that ranges from 3 to 32 nm depending on the ligand concentration. Protein recognition on the sensor surface is specific and has a concentration dependence that is similar to that in solution.
9
Translating Biomolecular Recognition into Nanomechanics
Jürgen Fritz, Marko Baller, H.P. Lang et al. · Science · 2000 · 1.6K citations
An Aptamer-Based Quartz Crystal Protein Biosensor
Michael A. Liss, Birgit Petersen, Hans Wolf et al. · Analytical Chemistry · 2002 · 507 citations
A femtojoule calorimeter using micromechanical sensors
J. R. Barnes, R. J. Stephenson, Charles N. Woodburn et al. · Review of Scientific Instruments · 1994 · 270 citations