Radiation induced current in the RF coils of integrated linac‐MR systems: The effect of buildup and magnetic field

15

Citations

11

References

2012

Year

Abstract

The RIC in RF coils results from the lack of electronic equilibrium in the coil conductor as the RIC in planar conductor was completely removed by identical buildup of adequate thickness to create electronic equilibrium. The buildup method of RIC removal is effective in cylindrical coil geometry when the coil conductor is in direct contact with the patient. The presence of air makes this method of RIC removal less effective although placing buildup still reduces the RIC by up to 60%. The RIC Monte Carlo simulation is a useful tool for practical coil design where radiation effects must be considered. The SNR is improved in the images obtained concurrently withradiation if buildup is applied to the coil.