ASAIO Journal · 2007 · 79 citations · 5 references
MedicineSurgeryCardiothoracic SurgeryTransplant SurgeryVentilator ManagementExtracorporeal CirculationExtracorporeal Life SupportExtracorporeal Membrane OxygenationLung TransplantationMechanical VentilationVentilatory SupportBlood FlowBridging To Lung TransplantationHypercapnic AcidosisMembrane OxygenationOxygenationRefractory HypoxemiaVentilator-associated HypoxemiaWeaning From Extracorporeal Membrane Oxygenation
Conventional extracorporeal membrane oxygenation and mechanical ventilation have both been identified as significant risk factors for post-lung transplant mortality when applied as a bridge to lung transplantation. We have previously described the successful use of the extracorporeal membrane ventilator Novalung as a bridge to lung transplantation in patients with severe hypercapnia and respiratory acidosis. In this setting the Novalung was connected in the arterio-venous mode without support of a mechanical blood pump. However, in patients with predominant hypoxemia, this pumpless mode does not achieve sufficient blood oxygenation due to limited blood flow. Thus, such patients require pump-driven support. Here we describe our initial experience with the Novalung extracorporeal membrane ventilator, which was originally designed for pumpless carbon dioxide removal, as a bridge to lung transplantation in patients with ventilator-refractory hypoxemia in the veno-venous pump-driven mode.
5
Elbert P. Trulock, Leah B. Edwards, David O. Taylor et al. · The Journal of Heart and Lung Transplantation · 2005 · 258 citations
Medicine, Cardiology, Surgery +14