Health ProfessionsOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsSpinal BiomechanicsSpine SurgeryCervical Spine BiomechanicsIntervertebral Disc BiomechanicsSpinal KinematicsSpinal KineticsSpinal Motion AnalysisSpinal Range Of MotionBiomechanical EffectsBiomechanical ResponseCervical Disc Degeneration MechanicsCervical KineticsCervical Motion Segment BiomechanicsCervical Spine MotionExperimental Cadaveric TestingMotion Preservation Biomechanics
STUDY DESIGN: An in vitro biomechanical study of C4-C5 intervertebral disc replacement using a cadaveric model. OBJECTIVES: To investigate the degree of motion afforded by a ball-and-socket cervical intervertebral disc prosthesis design. SUMMARY OF BACKGROUND DATA: Intervertebral disc prostheses designs attempt to restore or maintain cervical disc motion after anterior cervical discectomy and reduce the likelihood of accelerated degeneration in adjacent discs by maintaining normal motion at the affected disc level. Surprisingly, the actual kinetic and biomechanical effects that cervical disc arthroplasty imparts on the spine have not been widely reported. Accordingly, we investigated what effect implanting a cervical disc prosthesis has on the range of motion at the affected level as well as how it changes the coupled motion patterns at the level of implantation. METHODS: Six fresh-frozen human cadaveric cervical spines (C2-C7) were used in this study. We evaluated two different spinal conditions: intact and after disc replacement at C4-C5. Compression (using the follower load concept) and pure moment loading were applied to the specimen. Range of motion was measured using an optical tracking system. Statistical differences between the intact and replaced condition range of motion was determined using analysis of variance with post hoc comparisons (alpha = 0.05). RESULTS: The data indicate that the intervertebral disc prosthesis approximated the intact motion in all three rotation planes at the affected level. Finally, changes in cervical coupled rotations, specifically lateral bending during axial rotation loading and axial rotation during lateral bending loading, were not statistically significant between the two tested conditions. CONCLUSIONS: Our data demonstrate that a ball-and-socket design can replicate physiologic motion at the affected and adjacent levels. More importantly, the data indicate that motion coupling, which is most dramatic in the cervical spine and plays an important biomechanical role, is maintained.
13
Posterior C1–C2 Fusion With Polyaxial Screw and Rod Fixation
J. Harms, Robert P. Melcher · Spine · 2001 · 1.5K citations
Three-Dimensional Movements of the Upper Cervical Spine
Manohar M. Panjabi, Jiří Dvořák, JOANNE DURANCEAU et al. · Spine · 1988 · 334 citations
Static and Cyclical Biomechanical Analysis of Pedicle Screw Spinal Constructs
Bryan W. Cunningham, John C. Setter, Yasuhiro Shono et al. · Spine · 1993 · 185 citations