Stresses in cement mantles of hip replacements: effect of femoral implant sizes, body mass index and bone quality

J.M.S. Lamvohee, Rajshree Mootanah, P. Ingle, K. Cheah, John K. Dowell

Computer Methods in Biomechanics & Biomedical Engineering · 2009 · 15 citations · 21 references

DOIFull text

Open access

Concepts

Abstract

The effects of femoral prosthetic heads of diameters 22 and 28 mm were investigated on the stability of reconstructed hemi-pelves with cement mantles of thicknesses 1-4 mm and different bone qualities. Materialise medical imaging package and I-Deas finite element (FE) software were used to create accurate geometry of a hemi-pelvis from CT-scan images. Our FE results show an increase in cement mantle stresses associated with the larger femoral head. When a 22 mm femoral head is used on acetabulae of diameters 56 mm and above, the probability of survivorship can be increased by creating a cement mantle of at least 1 mm thick. However, when a 28 mm femoral head is used, a cement mantle thickness of at least 4 mm is needed. Poor bone quality resulted in an average 45% increase in the tensile stresses of the cement mantles, indicating resulting poor survivorship rate.