Analytical Chemistry · 1998 · 34 citations · 17 references
PharmacyChemistryAnalytical ChemistryElectroanalytical MethodsMass Spectrometric AnalysisBioanalytical Method ValidationLiquid ChromatographySample PreparationCapillary ElectrophoresisCapillary Electrophoresis Method DevelopmentElectrophoretic DetectionHydrodynamic InjectionIonic StrengthSample ConcentrationSample Matrix EffectsSample Preconcentration
An on-line sample concentration method using sample matrix switching and field amplification peak stacking has been developed. A microbore LC guard column is used to slightly retain the analytes in order to switch from a high ionic strength sample matrix (the physiological fluid) to a low ionic strength matrix (the LC mobile phase). The eluted LC peak is then trapped in a CE system and preconcentrated by field amplification peak stacking. The concentrated sample peak is then analyzed by CE. Compared to normal hydrodynamic injection, the sensitivity was increased by more than 500-fold without loss in resolution. A limit of detection of less than 10 nM for a physiological sample was achieved using UV adsorption detection. This method can be used for negatively or positively charged analytes.
17
Chao-Xuan Zhang, Wolfgang Thormann · Analytical Chemistry · 1998 · 146 citations
S. Pálmarsdóttir, Eddie Thordarson, L.‐E. Edholm et al. · Analytical Chemistry · 1997 · 146 citations