Radiology · 2000 · 244 citations · 26 references
MedicineCardiologyRadiologyDiagnostic ImagingDiagnostic RadiologyMedical PhysicsCardiac ImagingCardiac PacingCardiovascular ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingCardiac PacemakersIn Vivo StudiesIn-vitro CharacterizationIntracardiac PacingMagnetic ResonanceMagnetic Resonance PhysicsPacemaker ProgrammingPacing
PURPOSE: To evaluate the safety and feasibility of magnetic resonance (MR) imaging at 0.5 T in patients with implanted cardiac pacemakers. MATERIALS AND METHODS: Twenty-one models of pacemakers and 44 pacemaker electrodes were exposed to in vitro MR imaging with continuous registration of pacemaker output and temperature at the lead tip. Prior to MR imaging examination, pacemakers were programmed to an asynchronous mode (A00, V00, or D00). Pacemakers were examined before and after MR imaging. Forty-four patients with implanted pacemakers underwent 51 MR imaging examinations under cardiologic surveillance, continuous electrocardiography, pulse oximetry, and capnographic monitoring. RESULTS: MR imaging was safely performed in all patients. None of the pacemakers displayed a pacing dysfunction at MR imaging. No changes occurred in the programmed parameters in any device tested in vivo or in vitro. Maximum increases in the temperature at the lead tips were 8.90 degrees C at a specific absorption rate (SAR) of 0.6 W/kg and 23.50 degrees C under a worst-case radio-frequency (RF) heating condition with an SAR of 1.3 W/kg. CONCLUSION: MR imaging at 0.5 T can be safely performed in patients with implanted pacemakers in carefully selected clinical circumstances when appropriate strategies (programming to an asynchronous mode, adequate monitoring techniques, limited RF exposure) are used.
26
Recording of EEG during fMRI experiments: Patient safety
Louis Lemieux, Philip J. Allen, Florence Franconi et al. · Magnetic Resonance in Medicine · 1997 · 312 citations
Effects of magnetic resonance imaging on cardiac pacemakers and electrodes
Stephan Achenbach, Werner Moshage, B. Diem et al. · American Heart Journal · 1997 · 255 citations
Medicine, Cardiology, Radiology +14
The effects of nuclear magnetic resonance on patients with cardiac pacemakers.
William Pavlicek, Michael Geisinger, Lon Castle et al. · Radiology · 1983 · 208 citations
Medicine, Cardiology, Radiology +17
Temperature monitoring during radiofrequency catheter ablation of accessory pathways.
Jonathan J. Langberg, Hugh Calkins, R el-Atassi et al. · Circulation · 1992 · 179 citations · Full text