The Journal of the Acoustical Society of America · 1987 · 15 citations · 0 references
Natural SciencesPhysicsAcousticsUltrasonographyUltrasoundAcoustic MeasurementAcoustic ModelingPhysical AcousticsTransducer TechnologyUltrasonicsUltrasound PhysicsAcoustic EmissionAcoustic Transducer ModelingAngular DependenceTransducer CalibrationTransducer CharacterizationTransducersUltrasonic TransducersVector Calibration
The independent transient responses of a transducer to displacements polarized in each of the three orthogonal directions have been determined using a new method that assumes only linearity and no angular impingement dependence of the transducer response. The modeling of the transducer response is thus converted from a vector problem to three much more tractable scalar problems, thereby forming the basis for vector calibration of acoustic emission and ultrasonic transducers as receivers.