Inhibition of Biventricular Pacemakers by Oversensing of Far‐Field Atrial Depolarization

Igor Lipchenca, Stéphane Garrigue, Michael Glikson, S. Serge Barold, Jacques Clémenty

Pacing and Clinical Electrophysiology · 2002 · 38 citations · 0 references

Concepts

Abstract

LIPCHENCA, I., et al. : Inhibition of Biventricular Pacemakers by Oversensing of Far‐Field Atrial Depo‐larization. This report describes two patients who exhibited far‐field oversensing of the P wave by the ventricular channel of a DDD biventricular pacemaker implanted for the treatment of congestive heart failure. Oversensing probably occurred secondary to slight displacement of the left ventricular lead in the coronary venous system. Long‐term reliable pacing was restored by decreasing the sensitivity of the ventricular channel.