Review of Scientific Instruments · 1954 · 34 citations · 3 references
EngineeringNuclear EngineeringNuclear InstrumentationNuclear PhysicsRadiation DetectionDetector TechnologyGamma-ray DetectionGamma-ray DetectorsNeutron DetectionScintillation DetectorsScintillator MaterialsFast Neutron DetectionNeutron DetectorsNuclear RecoilPhotomultiplier DetectionPhotomultiplier DetectorsPhotomultiplier TubesRadiation Detection EfficiencyScintillation Neutron Detectors
A fast neutron scintillator adapted to end-window photomultipliers is described for neutron detection above ∼0.3-Mev energy. Light pulses are produced by the action of recoil protons from paraffine on zinc sulfide phosphor, the optimum weight ratio of the paraffine-zinc sulfide mixture being 3:2. Interfering gamma-ray pulses are suppressed by a 3 microsecond time-constant in the photomultiplier output circuit, and subsequent amplitude discrimination. A detection efficiency of 3.8 percent is attained for 4.1-Mev neutrons through the incorporation of ½-inch deep Lucite light guides.
3
W.F. Hornyak · Review of Scientific Instruments · 1952 · 86 citations
W. G. Moulton, Chalmers W. Sherwin · Review of Scientific Instruments · 1949 · 14 citations
A Scintillation Counter for Fast Neutrons
H. B. Frey · Review of Scientific Instruments · 1950 · 11 citations