Image information transfer properties of x‐ray intensifying screens in the energy range from 17 to 320 keV

A. Ginzburg, C. E. Dick

Medical Physics · 1993 · 47 citations · 0 references

Concepts

Abstract

The image information transfer properties of a number of x-ray fluorescent screens have been measured for x-ray energies from 17 to 320 keV. The detective quantum efficiency of the screens at each x-ray energy has been determined by separate measurements of the x-ray absorption efficiency and the statistical factor associated with the emission of optical photons upon absorption of an incident x-ray. Data have been recorded for both rare-earth phosphor screens and calcium tungstate screens. The value of the statistical factor for optical photon emission tends toward a constant value as the incident energy increases. Comparisons of the image information transfer properties are presented for several screens, which have been measured over a ten year interval. The utility of the screens for high-energy radiography is discussed.