Incidence of ‘Squeaking’ After Ceramic-on-Ceramic Total Hip Arthroplasty

Kenny Mai, Christopher A Verioti, Kace A. Ezzet, Steven N. Copp, Richard H. Walker, Clifford W. Colwell

Clinical Orthopaedics and Related Research · 2009 · 138 citations · 14 references

DOIFull text

Open access

Concepts

Abstract

UNLABELLED: The incidence of hip "squeak" associated with ceramic-on-ceramic bearings has been variably reported, ranging from 0.7% to 20.9%. We determined the patients' perception of squeaking in 306 patients (336 hips) in whom ceramic-on-ceramic total hip arthroplasties (THAs) were performed between 1997 and 2005. A questionnaire regarding hip noise was obtained by telephone. With a minimum followup of 2 years (mean, 3.9 years; range, 2-10 years), 290 patients (320 or 95% of the THAs) completed the questionnaire. Patients reported hip noise in 55 of the 320 THAs (17%); noise was perceived as squeak in 32 of the 320 (10%). Most squeaking hips (29 of 32) were pain-free and symptom-free. One patient was unhappy with his squeaking hip without pain. Our data suggest a much higher incidence of squeak as perceived by patients than previously reported. LEVEL OF EVIDENCE: Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

References