2013 · 59 citations · 10 references
EngineeringBioengineeringBiomedical EngineeringBiomedical Signal ProcessingBiosensorsSensorsCardiovascular MonitoringElectrocardiographyPressure SensorsVital SignsVital Signs RecordingWearable BiosensorsBlood PressureBlood Pressure EstimationBlood Pressure MeasurementBlood Pressure MonitoringElectrocardiogram Blood Pressure EstimationPulse Wave PropagationWearable Vital Signs MonitoringWearable Vital Signs Sensors
Recently, monitoring of blood pressure fluctuation in the daily life is focused on in the hypertension care area to predict the risk of cardiovascular and cerebrovascular disease events. In this paper, in order to propose an alternative system to the existed ambulatory blood pressure monitoring (ABPM) sphygmomanometer, we have developed a prototype of small wearable device consisting of electrocardiogram (ECG) and photopelthysmograph (PPG) sensors. In addition, it was examined whether blood pressure can be estimated based on pulse wave transit time (PWTT) only by attaching that device on the surface of the chest. We indicated that our system could also sense tendency of time-dependent change of blood pressure by measuring pulse of vessel over the sternum while its propagation distance is short.
10
Continuous noninvasive blood pressure measurement by pulse transit time
Parry Fung, Guy A. Dumont, Craig R. Ries et al. · 2005 · 174 citations
Medicine, Cardiology, Anesthesia +15