Six-axis multi-anvil press for high-pressure, high-temperature neutron diffraction experiments

Asami Sano‐Furukawa, Takanori Hattori, Hiroshi Arima, Akihiro Yamada, Satoshi TABATA, Mutsumi Kondo, Akihiro Nakamura, Hiroyuki Kagi, T. Yagi

Review of Scientific Instruments · 2014 · 80 citations · 14 references

Concepts

Abstract

We developed a six-axis multi-anvil press, ATSUHIME, for high-pressure and high-temperature in situ time-of-flight neutron powder diffraction experiments. The press has six orthogonally oriented hydraulic rams that operate individually to compress a cubic sample assembly. Experiments indicate that the press can generate pressures up to 9.3 GPa and temperatures up to 2000 K using a 6-6-type cell assembly, with available sample volume of about 50 mm(3). Using a 6-8-type cell assembly, the available conditions expand to 16 GPa and 1273 K. Because the six-axis press has no guide blocks, there is sufficient space around the sample to use the aperture for diffraction and place an incident slit, radial collimators, and a neutron imaging camera close to the sample. Combination of the six-axis press and the collimation devices realized high-quality diffraction pattern with no contamination from the heater or the sample container surrounding the sample. This press constitutes a new tool for using neutron diffraction to study the structures of crystals and liquids under high pressures and temperatures.

References