A slit method to determine the focal spot size and shape of TomoTherapy system

13

Citations

17

References

2011

Year

Abstract

Our study of the focal spot measurement using slit-collimators showed that accurate source profile measurements can be achieved through fitting of measurement results at different slit widths and source-to-slit distances (SSD). Quantitative measurements of the TomoTherapy linac focal spot showed that the source distribution could be better described with a model consisting of two Gaussian components rather than a single Gaussian model as assumed in previous studies.