Publication | Open Access
Positive corona discharge ion source with IMS/MS to detect impurities in high purity Nitrogen
19
Citations
18
References
2011
Year
Chemical EngineeringElectrical EngineeringEngineeringGlow DischargeMass SpectrometryCorona DischargePositive Corona DischargeAnalytical ChemistryHydrogenInstrumentationChemistryGas Discharge PlasmaHigh Purity NitrogenIon ProcessIon EmissionIon Mobility
We have applied the Ion Mobility Spectrometry/Mass Spectrometry (IMS/MS) and the Atmospheric Pressure Chemical Ionisation/Mass Spectrometry (APCI/MS) techniques to study the formation of the ions in the positive corona discharge (CD) in highly purified nitrogen with impurities at 100 ppt level. The main products observed were H3O+(H2O)n ions (reduced ion mobility of 2.15 cm2 V-1 s-1). Additionally, we have observed ions with reduced mobilities 2.42 cm2 V-1 s-1 and 2.30 cm2 V-1 s-1. The intensity of these ions was increasing with the increasing discharge current. We associated these peaks with NH4+ and NO+(H2O)n. The formation of these ions results from trace amounts of O2 and NH3 in nitrogen. The time evolution of the ions in corona discharge has been studied using the APCI/MS technique in the time windows ranging from 100 μs to 2 ms. The present work indicates the ability of the IMS technique equipped with CD ions source to detect impurities below 100 ppt level.
| Year | Citations | |
|---|---|---|
Page 1
Page 1