Real-time detection of pneumothorax using electrical impedance tomography*

Eduardo Leite Vieira Costa, Caroline N. Chaves, Susimeire Gomes, Marcelo A. Beraldo, Márcia Souza Volpe, Mauro R. Tucci, Ivany A. L. Schettino, Stephan H. Böhm, Carlos Roberto Ribeiro de Carvalho, Harki Tanaka,

Critical Care Medicine · 2008 · 195 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVES: Pneumothorax is a frequent complication during mechanical ventilation. Electrical impedance tomography (EIT) is a noninvasive tool that allows real-time imaging of regional ventilation. The purpose of this study was to 1) identify characteristic changes in the EIT signals associated with pneumothoraces; 2) develop and fine-tune an algorithm for their automatic detection; and 3) prospectively evaluate this algorithm for its sensitivity and specificity in detecting pneumothoraces in real time. DESIGN: Prospective controlled laboratory animal investigation. SETTING: Experimental Pulmonology Laboratory of the University of São Paulo. SUBJECTS: Thirty-nine anesthetized mechanically ventilated supine pigs (31.0 +/- 3.2 kg, mean +/- SD). INTERVENTIONS: In a first group of 18 animals monitored by EIT, we either injected progressive amounts of air (from 20 to 500 mL) through chest tubes or applied large positive end-expiratory pressure (PEEP) increments to simulate extreme lung overdistension. This first data set was used to calibrate an EIT-based pneumothorax detection algorithm. Subsequently, we evaluated the real-time performance of the detection algorithm in 21 additional animals (with normal or preinjured lungs), submitted to multiple ventilatory interventions or traumatic punctures of the lung. MEASUREMENTS AND MAIN RESULTS: Primary EIT relative images were acquired online (50 images/sec) and processed according to a few imaging-analysis routines running automatically and in parallel. Pneumothoraces as small as 20 mL could be detected with a sensitivity of 100% and specificity 95% and could be easily distinguished from parenchymal overdistension induced by PEEP or recruiting maneuvers. Their location was correctly identified in all cases, with a total delay of only three respiratory cycles. CONCLUSIONS: We created an EIT-based algorithm capable of detecting early signs of pneumothoraces in high-risk situations, which also identifies its location. It requires that the pneumothorax occurs or enlarges at least minimally during the monitoring period. Such detection was operator-free and in quasi real-time, opening opportunities for improving patient safety during mechanical ventilation.

References

19