Publication | Closed Access
Absolute measurement of detector quantum efficiency using parametric downconversion
116
Citations
9
References
1987
Year
Silicon Avalanche PhotodiodeQuantum SciencePhotonicsQuantum PhotonicsEngineeringQuantum ComputingPhysicsNatural SciencesApplied PhysicsQuantum MeasurementQuantum InformationPhotonic Integrated CircuitQuantum EntanglementQuantum Photonic DeviceOptoelectronicsDetector Quantum EfficiencyAbsolute Quantum EfficiencyParametric Downconversion Crystal
We show that a parametric downconversion crystal emitting angle resolved coincident photon pairs can be used to measure the absolute quantum efficiency of a photon counting detection system. We have measured the quantum efficiency of a silicon avalanche photodiode, operated in Geiger mode, as a function of operating voltage and compare this to results obtained using a conventional method.
| Year | Citations | |
|---|---|---|
1970 | 1.1K | |
1975 | 624 | |
1985 | 197 | |
1986 | 185 | |
1987 | 127 | |
1987 | 123 | |
1977 | 88 | |
1986 | 78 | |
1985 | 61 |