Pulmonary Vein Stenosis After Catheter Ablation of Atrial Fibrillation

Ivan M. Robbins, Edward V. Colvin, Thomas P. Doyle, W. Evans Kemp, James E. Loyd, William McMahon, G. Neal Kay

Circulation · 1998 · 447 citations · 13 references

Concepts

Abstract

BACKGROUND: This report describes the complication of pulmonary vein stenosis with resultant severe pulmonary hypertension that developed in 2 patients after successful catheter ablation of chronic atrial fibrillation. METHODS AND RESULTS: Three months after successful catheter ablation of atrial fibrillation, both patients developed progressive dyspnea and pulmonary hypertension. Both were found to have severe stenosis of all 4 pulmonary veins near the junction with the left atrium. Balloon dilation of the stenotic pulmonary veins was performed in these patients, with improvement in dyspnea and pulmonary hypertension. CONCLUSIONS: The complication of pulmonary vein stenosis is potentially life-threatening, and the application of radiofrequency current within the pulmonary veins with standard catheter technology should be avoided. This complication can be treated with balloon dilation, although the long-term course is unknown.

References

13