Validation of Volume Blood Flow Measurements Using Three-Dimensional Distance-Concentration Functions Derived from Digital X-Ray Angiograms

David J. Hawkes, Alexander M. Seifalian, A. C. F. Colchester, Najma Iqbal, Charles R. Hardingham, Christopher F. Bladin, Kenneth E.F. Hobbs

Investigative Radiology · 1994 · 25 citations · 0 references

Concepts

Abstract

RATIONALE AND OBJECTIVES: The authors present phantom validation of a method for computing pulsatile flow waveforms in arterial vessels from high-frame-rate biplane x-ray angiograms. METHODS: The three-dimensional course of a blood vessel is constructed from biplane digital x-ray angiograms. A parametric image of contrast mass versus time and true three-dimensional path length is generated. Adjacent contrast mass-distance profiles are matched to compute instantaneous velocity, which is multiplied by cross-sectional area to yield volume flow. An electromagnetic flowmeter was used to validate flow estimates in a phantom consisting of 150-mm tubes 3, 4, and 6 mm in diameter, orientated 15 degrees, 30 degrees, and 35 degrees to the imaging plane, with flow rates and waveforms expected in vivo. RESULTS: Mean and peak flows were accurate to within 9% and 10%, respectively, for velocities of less than 1 meter/second at a frame rate of 25 frames per second. CONCLUSIONS: A practical method for computing highly pulsatile flow waveforms in vivo in tortuous vessels is presented.